The Guardian SRM-232 is a handheld sheet resistivity measurement system consisting of a portable meter and a four-point probe head for accurately measuring the sheet resistance of conductive coatings and thin-film materials. Typical applications include conductive paints, EMI shielding coatings, indium tin oxide (ITO) coatings on glass, and other conductive surface materials.

Powered by a single 9-volt battery, the SRM-232 provides portable measurement capability for laboratory, production, and field environments.

Surface Resistivity Meter

The Guardian SRM-232 is a handheld sheet resistivity measurement system consisting of a portable meter and a four-point probe head for accurately measuring the sheet resistance of conductive coatings and thin-film materials. Typical applications include conductive paints, EMI shielding coatings, indium tin oxide (ITO) coatings on glass, and other conductive surface materials.

Powered by a single 9-volt battery, the SRM-232 provides portable measurement capability for laboratory, production, and field environments.

KIT CONFIGURATION

  • Surface Resistivity Meter
  • Four-Point Probe Heads
  • RS-232 Interface Cable
  • RJ45 Ethernet Coil Cable
  • SRM Calibration Fixture

  • Protective Pelican Case

Kit Configuration

  • Surface Resistivity Meter
  • Four-Point Probe Heads
  • RS-232 Interface Cable
  • RJ45 Ethernet Coil Cable
  • SRM Calibration Fixture

  • Protective Pelican Case

Instrument Options

The Guardian SRM-232 is available in four measurement ranges to accommodate conductive materials with different sheet-resistance levels. This allows users to select the range best suited for applications such as conductive coatings, EMI shielding materials, ITO-coated glass, thin films, and solar-cell wafers. Each model provides a range tailored to the resistance level of the material being tested.

Each new instrument includes a NIST-traceable certificate of calibration and one-year factory warranty.

  • SRM-232-10

    Range: 0.000 to 9.999 Ω/sq
    Resolution: 0.004 Ω/sq
    Accuracy: 0.01 Ω/sq at 1.00 Ω/sq

  • SRM-232-1000

    Range: 0.00 to 999.99 Ω/sq
    Resolution: 0.40 Ω/sq
    Accuracy: 0.70 Ω/sq at 100 Ω/sq

  • SRM-232-100

    Range: 0.00 to 95.00 Ω/sq
    Resolution: 0.04 Ω/sq
    Accuracy: 0.07 Ω/sq at 10 Ω/sq

  • SRM-232-2000

    Range: 0.0 to 2000.00 Ω/sq
    Resolution: 0.80 Ω/sq
    Accuracy: 1.4 Ω/sq at 1000 Ω/sq

Probe Head Options

The Guardian SRM Probe Head is a four-point probe assembly with a durable plastic housing designed for direct contact measurements. The probe connects to the SRM-232 unit through a RJ45 Ethernet cable. A flush-mount housing option is available that recesses the probe tips within the housing to limit tip travel and reduce applied spring force.

  • Tip Spacing

    40 mils | 50 mils | 62.5 mils

  • Spring Load

    45g | 85g | 180g per tip

  • Tip Radii

    1.6 mils | 5 mils | 10 mils

  • Tip Material

    Tungsten Carbide

Calibration Fixture

The Guardian Calibration Fixture is designed to verify the accuracy of the SRM-232. The fixture contains three precision reference points across the SRM-232 measurement range: a dead short, a precision resistor representing the middle of the measurement range, and a precision resistor near the upper end of the range.

  • The SRM-CAL allows users to verify the SRM-232 instrument performance against an external standard.

  • Each Calibration Fixture is supplied with an NIST-traceable certificate of calibration and can be recertified annually to support ongoing measurement verification requirements.

  • The Calibration Fixture does not provide end-user adjustment capability for the SRM-232 instrument. If an SRM-232 requires adjustment, a factory recalibration service must be performed using specialized equipment.

Instrument Options

The Guardian SRM-232 is available in four measurement ranges to accommodate conductive materials with different sheet-resistance levels. This allows users to select the range best suited for applications such as conductive coatings, EMI shielding materials, ITO-coated glass, thin films, and solar-cell wafers. Each model provides a range tailored to the resistance level of the material being tested.

Each new instrument includes a NIST-traceable certificate of calibration and one-year factory warranty.

  • SRM-232-10

    Range: 0.000 to 9.999 Ω/sq
    Resolution: 0.004 Ω/sq
    Accuracy: 0.01 Ω/sq at 1.00 Ω/sq

  • SRM-232-100

    Range: 0.00 to 95.00 Ω/sq
    Resolution: 0.04 Ω/sq
    Accuracy: 0.07 Ω/sq at 10 Ω/sq

  • SRM-232-1000

    Range: 0.00 to 999.99 Ω/sq
    Resolution: 0.40 Ω/sq
    Accuracy: 0.70 Ω/sq at 100 Ω/sq

  • SRM-232-2000

    Range: 0.0 to 2000.00 Ω/sq
    Resolution: 0.80 Ω/sq
    Accuracy: 1.4 Ω/sq at 1000 Ω/sq

Probe Head Options

The Guardian SRM Probe Head is a four-point probe assembly with a durable plastic housing designed for direct contact measurements. The probe connects to the SRM-232 unit through a RJ45 Ethernet cable. A flush-mount housing option is available that recesses the probe tips within the housing to limit tip travel and reduce applied spring force.

  • Tip Spacing

    40 mils | 50 mils | 62.5 mils

  • Spring Load

    45g | 85g | 180g per tip

  • Tip Radii

    1.6 mils | 5 mils | 10 mils

  • Tip Material

    Tungsten Carbide

Calibration Fixture

The Guardian Calibration Fixture is designed to verify the accuracy of the SRM-232. The fixture contains three precision reference points across the SRM-232 measurement range: a dead short, a precision resistor representing the middle of the measurement range, and a precision resistor near the upper end of the range.

  • The SRM-CAL allows users to verify the SRM-232 instrument performance against an external standard.

  • Each Calibration Fixture is supplied with an NIST-traceable certificate of calibration and can be recertified annually to support ongoing measurement verification requirements.

  • The Calibration Fixture does not provide end-user adjustment capability for the SRM-232 instrument. If an SRM-232 requires adjustment, a factory recalibration service must be performed using specialized equipment.

Kit Configuration

  • Surface Resistivity Meter
  • Four-Point Probe Heads
  • RS-232 Interface Cable
  • RJ45 Ethernet Coil Cable
  • SRM Calibration Fixture

  • Protective Pelican Case

Instrument Options

The Guardian SRM-232 is available in four measurement ranges to accommodate conductive materials with different sheet-resistance levels. This allows users to select the range best suited for applications such as conductive coatings, EMI shielding materials, ITO-coated glass, thin films, and solar-cell wafers. Each model provides a range tailored to the resistance level of the material being tested.

Each new instrument includes a NIST-traceable certificate of calibration and one-year factory warranty.

  • SRM-232-10

    Range: 0.000 to 9.999 Ω/sq
    Resolution: 0.004 Ω/sq
    Accuracy: 0.01 Ω/sq at 1.00 Ω/sq

  • SRM-232-1000

    Range: 0.00 to 999.99 Ω/sq
    Resolution: 0.40 Ω/sq
    Accuracy: 0.70 Ω/sq at 100 Ω/sq

  • SRM-232-100

    Range: 0.00 to 95.00 Ω/sq
    Resolution: 0.04 Ω/sq
    Accuracy: 0.07 Ω/sq at 10 Ω/sq

  • SRM-232-2000

    Range: 0.0 to 2000.00 Ω/sq
    Resolution: 0.80 Ω/sq
    Accuracy: 1.4 Ω/sq at 1000 Ω/sq

Probe Head Options

The Guardian SRM Probe Head is a four-point probe assembly with a durable plastic housing designed for direct contact measurements. The probe connects to the SRM-232 unit through a RJ45 Ethernet cable. A flush-mount housing option is available that recesses the probe tips within the housing to limit tip travel and reduce applied spring force.

  • Tip Spacing

    40 mils | 50 mils | 62.5 mils

  • Spring Load

    45g | 85g | 180g per tip

  • Tip Radii

    1.6 mils | 5 mils | 10 mils

  • Tip Material

    Tungsten Carbide

Calibration Fixture

The Guardian Calibration Fixture is designed to verify the accuracy of the SRM-232. The fixture contains three precision reference points across the SRM-232 measurement range: a dead short, a precision resistor representing the middle of the measurement range, and a precision resistor near the upper end of the range.

  • The SRM-CAL allows users to verify the SRM-232 instrument performance against an external standard.

  • Each Calibration Fixture is supplied with an NIST-traceable certificate of calibration and can be recertified annually to support ongoing measurement verification requirements.

  • The Calibration Fixture does not provide end-user adjustment capability for the SRM-232 instrument. If an SRM-232 requires adjustment, a factory recalibration service must be performed using specialized equipment.

YOU’RE IN GOOD COMPANY

We’re ready to support your project

YOU’RE IN GOOD COMPANY

We’re ready to support your project

We’re ready to support your project